To minimize the delivered concentration of NO
WARNING:
for use with the manual resuscitator bags:
• Use the smallest bag adequate to deliver the desired tidal volume.
• Oxygen tubing lengths greater than 72 inches should not be used (between the
injector module and the bag).
• Use the highest fresh gas flow rated (up to 15 L/min) that is practical.
• Use the lowest practical inspired oxygen concentration.
• After starting fresh gas flow, squeeze the bag several times to empty residual
gas in the bag prior to using the system to ventilate a patient.
3
2
1
13
1. O
Flowmeter
2
2. Injector Module Electrical Cable
3. NO/N
Injector Tube
2
4. Patient Gas Sample Line with Nafion
5. O
Tubing
2
6. O
Tubing Sample Tee
2
7. Hyper-Inflation Bag
Figure 3-23 Example: Hyperinflation Manual Bagging System Diagram
Testing has been conducted using the following hyperinflation and self-inflating bag systems.
• Hudson RCI Hyperinflation 1 L Adult # 5404
• Hudson RCI Hyperinflation 0.5 L Neonatal # 5403
• Nellcor-Puritan Bennett Self-inflating 1.76 L Adult # 655005
• Nellcor-Puritan Bennett Self-inflating 0.52 L Infant # 616416
3-30
12
11
10
9
, the following steps should be taken
2
8
8. Pressure Gauge
9. 15M X 4.5 mm Adapter
10. Injector Module
11. 22M/15F X 22M/15F Adapter
12. 15M X 4.5 mm Adapter
13. O
Tubing
2
4
5
6
7
Part No. 20717 Rev-01
2014-07