Mallinckrodt INOMAX DSir Plus Operation Manual page 145

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

Alarm
24. Water Trap Bottle Full
25. Low Calibration Failed
Part No. 20717 Rev-01
2014-07
Low Priority Alarms
Possible Cause
a. The water trap bottle on the
side of the INOmax DS
IR
full.
b. Water trap bottle is empty
but the message remains in
the alarm message box.
a. Zeroing valve has failed to
switch.
b. Sampling patient circuit
gases during the low
calibration.
Recommended Action
All of these actions can be performed while
delivering INOMAX to the patient:
Empty water trap bottle.
is
Remove water trap bottle and clean optical
sensor level indicator with an alcohol swab.
Contact Technical Support.
All of these actions can be performed while
delivering INOMAX to the patient:
Repeat low calibration.
Wait for low calibration to complete
(approximately three minutes).
Perform a manual low calibration.
Contact Technical Support.
5-15