Inoblender Circuit Connection - Mallinckrodt INOmax DSIR Pocket Manual

Delivery system, series 3 software
Hide thumbs Also See for INOmax DSIR:
Table of Contents

Advertisement

INOblender Circuit
Connection Diagram
INOblender Warnings:
• The purge procedure must be followed to help ensure NO
purged from the pressure regulator, INOblender and hoses
before the manual resuscitator bag is connected to the patient.
The manual bag should be squeezed continuously during use
to avoid NO
continuously while delivering INOMAX, the bag should be
removed from the patient and the purge procedure performed
before continuing.
• Persons using this device should be trained on and
experienced in the use of this device to assure effective
administration of INOMAX and to avoid injury to the patient or
others resulting from inhalation of excess INOMAX, nitrogen
dioxide or other reaction products.
INOblender Cautions:
• Refer to the manufacturer's procedures for using the resuscitation
bag. When finished, turn the INOMAX cylinder off and continue to
flow O
until the NO pressure gauge reads zero, then turn the O
2
flow off and the NO dial to zero ppm.
Note: Connections to various ventilators as well as their corresponding
disposable circuits, are unique to each manufacturer. Please
refer to the specific breathing device operation manual or
instructions for use for guidance.
Part No. 20751 Rev-01
2014-08
building up in the bag. If the bag is not squeezed
2
50
is
2
2