Duration Chart; Inomax Cylinder D-Size - Mallinckrodt INOMAX DSir Plus Operation Manual

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

Duration Chart

INOMAX Cylinder D-Size

For an D-Size 800 ppm Cylinder Concentration* (typically used in transport)
(Illustrative Only)
5 L/min
5
7.0 Days
10
3.5 Days
20
1.7 Days
40
20 Hours
80
9.5 Hours
Typically used in transport
This chart is representative of a range of doses available on the INOmax DS
are not the recommended therapeutic dose.
* All calculations for the table above are based on a full cylinder of 138 bar (2000 psig), 353 liter "D" cylinder,
and also accounting for cylinder change at 14 bar (200 psig). The figures are calculated on total continuous
flow cylinder conversion factor (2.6 liters per bar and 0.18 liters per psig).
• INOMAX flow = [Desired dose X total ventilator flow] ÷ [Cylinder concentration - desired dose]
• Cylinder volume = Cylinder conversion factor X cylinder pressure (bar/psig)
• Cylinder duration = Cylinder volume ÷ INOMAX flow rate
Calculations are considered estimates and may vary under clinical circumstances.
For more information, call 1-877-KNOW-INO (1-877-566-9466)
4-12
FLOW
10 L/min
20 L/min
3.5 Days
1.8 Days
1.7 Days
21 Hours 10.5 Hours
20.7 Hours 10.3 Hours 5.2 Hours
10 Hours
5 Hours
4.8 Hours
2.4 Hours
40 L/min
21 Hours
2.5 Hours
1.2 Hours
and doses higher than 20 ppm
IR