Mallinckrodt INOMAX DSir Plus Operation Manual page 141

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

Alarm
10. Drug Past Expiry Date
Drug Concentration
11.
Mismatch
12. Injector Module Fail
Part No. 20717 Rev-01
2014-07
High Priority Alarms
Possible Cause
a. INOMAX cylinder is
expired.
Note: The INOmax DS
IR
recognizes the drug as
expired on the first day of
labeled expiration month on
the INOMAX cylinder.
a. INOMAX cylinder is the
wrong concentration.
a. The Injector Module
electrical cable may be
disconnected.
b. The Injector Module may
have failed.
c. The Injector Module
electrical cable may have
failed.
Recommended Action
If dose is set, "Delivery Stopped" will occur
two minutes from point when cylinder valve is
opened.
1. Close expired cylinder valve.
2. Remove expired INOMAX cylinder from
INOmax DS
cart.
IR
3. Connect an INOMAX cylinder with a valid
expiration date.
Contact Technical Support.
If dose is set, "Delivery Stopped" will occur
two minutes from point when cylinder valve is
opened.
1. Close mismatched cylinder valve.
2. Remove wrong concentration INOMAX
cylinder from INOmax DS
3. Connect an INOMAX cylinder to
INOmax DS
with a valid concentration.
IR
Contact Technical Support.
INOMAX delivery has been interrupted.
Manually ventilate patient with the INOblender
or
turn integrated pneumatic backup delivery ON
(verify at least five L/min ventilator gas flow in
patient circuit).
Reconnect injector module cable.
Change injector module. Pull back on knurled
sleeve to release connector.
Change injector module cable.
Contact Technical Support.
cart.
IR
5-11