(Delivery System) Pocket Guide Series 3 software Automated Pre-Use Procedure Integrated Pneumatic Backup INOMAX Delivery ® Transport Regulator/Cap Assembly Oxygen Dilution Chart INOMAX Cylinder Duration Chart Circuit Connection Diagrams Changing INOMAX Cylinders High Calibration Connection Diagrams INOmax DS Disposable Adapters For 24 Hour Assistance Call 1-877-566-9466 Part No.
Page 2
IMPORTANT: This guide is provided as a convenience and for general information only. Do not use this product without clearly and thoroughly understanding the most recent revision of the INOmax DS Operation Manual. The Operation ® Manual is the source for specific, updated information regarding warnings, cautions, checklists, diagrams, and/or instructions contained in this guide.
Pre-Use Checkout Automated Pre-Use Checkout 1. Turn INOmax DS ON, verify speaker function. Note: A low range calibration automatically starts following the self test. A Pre-Use wizard will be displayed on the main screen, which will provide step-by-step instructions to complete the automated Pre-Use procedure.
Page 6
Pre-Use Checkout Part No. 20751 Rev-01 2014-08...
Page 7
Pre-Use Checkout (Intentionally left blank) Part No. 20751 Rev-01 2014-08...
Integrated Pneumatic Backup INOMAX Delivery Note: Use the integrated pneumatic backup function only for a short time, until a replacement delivery system can be obtained. The INOblender can also be used as a backup. If the main delivery system fails, the INOmax DS has an integrated pneumatic backup delivery function that allows the patient to remain connected to the ventilator.
Page 9
Integrated Pneumatic Backup INOMAX Delivery The main screen: · Indicates that backup delivery is on and that the set dose is turned off. · Displays the estimated dose that the patient should be receiving, based on the ventilator flow. · Displays the NO concentration table Note: If the injector module is not functioning, the estimated backup dose graphic will be inactive.
Transport Regulator/ Cap Assembly Transport Regulator/Cap Assembly WARNING: • A new INOMAX cylinder and regulator must be purged before use to ensure the patient does not receive an excess level of NO • Loss of communication between the INOmax DS and the INOMAX cylinder for more than one hour will result in interruption of INOMAX delivery.
Page 11
Transport Regulator/ Cap Assembly Step Two Connect the INOMAX regulator hose to one of the INOMAX inlets on the back of the INOmax (see Figure 3). Figure 3 Step Three Connect the Infrared cable from the Transport Regulator/Cap Assembly to the back of the INOmax (see Figure 4).
Page 12
Transport Regulator/ Cap Assembly Step Four Place the Cap Assembly over the INOmeter (see Figure 5). Note: Be sure to align the keyway inside the Cap Assembly with the iButton on the INOmeter (see Figure 5 and 6). iButton The electrical cord exits the cap directly above the...
Page 13
Transport Regulator/ Cap Assembly Final Set-up Diagram The following diagram and photo illustrates all of the components connected. Additional Information Communication will take place between the INOmax DS and the INOmeter after the boot up phase of the INOmax DS is complete.
Duration Chart (88-size) INOMAX Cylinder 88-Size For an 88-Size 800 ppm Cylinder Concentration* (Illustrative Only) FLOW 5 L/min 10 L/min 20 L/min 40 L/min 39 Days 19.5 Days 9.8 Days 4.9 Days 19.4 Days 9.7 Days 4.8 Days 2.4 Days 9.6 Days 4.8 Days 2.4 Days...
Circuit Connection Diagrams Proper use of these products depends on careful reading and understanding of labeling and instructions. Please refer to the INOmax DS and INOblender operation manuals for guidance. Also refer to the specific breathing device operation manual or instructions for use. INOmax DS Warnings: • INOmax DS subtracts gas from the breathing circuit via the gas sampling system at 230 mL per minute;...
Circuit Connection Diagrams Acutronics Medical Systems AG Fabian +nCPAP Evolution (validated for use outside of the United States) 1. Fabian+ nCPAP Evolution 8. 22F X 15M Adapter 2. Patient Gas Sample Line with Nafion 9. Humidifier 3. Injector Module Electrical Cable 10. Inspiratory Breathing Circuit Hose 4.
Circuit Connection Diagrams Acutronics Medical Systems AG Fabian HFO (validated for use outside of the United States) 1. Fabian HFO Ventilator 9. 22F X 15M Adapter 2. Patient Gas Sample Line with Nafion 10. Humidifier 3. Injector Module Electrical Cable 11. Inspiratory Breathing Circuit Hose 4.
Circuit Connection Diagrams Bagging Systems While Using the Injector Module WARNING: To minimize the delivered concentration of NO2, the following steps should be taken for use with the manual resuscitator bags: • Use the smallest bag adequate to deliver the desired tidal volume.
Page 23
Circuit Connection Diagrams 1. O 2 Flowmeter (wall outlet or cylinder) 2. O 2 Tubing 3. 15M X 4.5 mm Adapter 4. 22M/15F X 22M/15F Adapter 5. Injector Module 6. 15M X 4.5 mm Adapter 7. O 2 Tubing 8. O 2 Tubing Sample Tee 9. Patient Gas Sample Line with Nafion 10.
Page 24
• The monitoring system within the INOmax DSIR will not detect generation of within the hyperinflation bag or self-inflating bag devices and the alarms...
Page 25
Circuit Connection Diagrams Bagging Systems While Using the Injector Module 1. O Flowmeter 8. Pressure Gauge 2. Injector Module Electrical Cable 9. 15M X 4.5mm Adapter 3. NO/N Injector Tube 10. Injector Module 4. Patient Gas Sample Line with Nafion 11. 22M/15F X 22M/15F Adapter 5.
Circuit Connection Diagrams Bunnell Life Pulse High Frequency Ventilator Circuit WARNING: • The INOmax DS backup mode (250 mL/min.) should not be used with the Bunnell Life Pulse as ventilator flow rates are normally below the recommended ventilator flows. • Place the Life Pulse in Standby prior to suctioning the patient to avoid NO delivery transiently exceeding the set dose by up to 30 ppm.
Page 27
Circuit Connection Diagrams Bunnell Life Pulse High Frequency Ventilator Circuit (cont.) 1. INOmax DS 8. Sample Tee 2. Bunnell Life Pulse 9. Patient Box 3. Humidifier 10. One-Way Valve 4. Humidifier 11. Injector Module 5. Conventional Ventilator 12. NO/N Injector Tube 6.
Circuit Connection Diagrams Connecting INOmax DS Sample Tee to the Bunnell Life Pulse Circuit 1. From Patient Box 2. Cut Green tube at midpoint (approximately six inches from the Life Port Adapter) 3. Patient Gas Sample Line with Nafion 4. Insert Sample Tee 5.
Circuit Connection Diagrams Circle Anesthesia System 1. Patient Gas Sample Line with 8. Absorber Inspiratory Port Nafion 9. Absorber 2. Patient Gas Sample Line 10. Injector Module Input Connection a. Injector Module Input End 3. INOmax DS b. Injector Module Output End 4.
Circuit Connection Diagrams Fisher & Paykel Optiflow Breathing Circuit 1. Patient Gas Sample Line with Nafion 10. Breathing Circuit 2. INOmax DS 11. Temperature Probe 3. Oxygen Source 12. Gas Sample Tee 4. Breathing Circuit Hose 13. 22M/15F X 22M/15F Adapter 5. Injector Module 14. 22 mm ID X 22 mm ID Cuff Adapter 6.
Circuit Connection Diagrams Sensormedics 3100A/B High Frequency Oscillatory Ventilator with a Filtered Circuit a. Injector Module b. 22F Inlet c. 22M / 15F Outlet Part No. 20751 Rev-01 2014-08...
Circuit Connection Diagrams Sensormedics 3100A/B High Frequency Oscillatory Ventilator with a Rigid or Flexible Circuit a. Injector Module b. 22F Inlet c. 22M / 15F Outlet 9. Humidifier Inlet 1. Sensormedics 3100A/B Ventilator 10. Humidifier Outlet 2. Ventilator Outlet 11. Patient Gas Sample 3.
Circuit Connection Diagrams SLE Life Support SLE5000 Note: • Validated for use outside of the United States. • A one-way valve is not required for use during high frequency ventilation mode. 1. Patient Wye 7. Injector Module 2. Patient Gas Sample Line with Nafion 8. 22F X 15M Adapter 3.
Circuit Connection Diagrams Spontaneous Breathing Patient on a Mask Circuit 1. O Tubing 10. One-Way Valve 2. 15M X 4.5 mm Adapter 11. Sealed Face Mask 3. 22M/15F X 22M/15F Adapter 12. One-Way Valve 4. Breathing Circuit Tee 13. Patient Gas Sample Line with Nafion 5.
Circuit Connection Diagrams Spontaneous Breathing Patient on a Nasal Cannula The INOmax DS can be used with nasal cannula to deliver INOMAX concentrations from 5-80 ppm and an oxygen flow rate as low as 2 L/min. WARNING: Do not use the INOmax DS backup mode with flow rates less than 5 L/min. 1.
• Follow all manufacturer instructions for connection to the Vapotherm Precision Flow. 1. Patient Gas Sample Line with Nafion 2. INOmax DSIR 3. Precision Flow Unit 4. Injector Module 5. Patient Delivery Tube 6. Oxygen Tubing Sample Tee 7. Patient Cannula 8.
INOblender Circuit Connection Diagram INOblender Warnings: • The purge procedure must be followed to help ensure NO purged from the pressure regulator, INOblender and hoses before the manual resuscitator bag is connected to the patient. The manual bag should be squeezed continuously during use to avoid NO building up in the bag.
Changing INOMAX Cylinders Changing INOMAX Cylinders WARNING: • A new INOMAX cylinder and regulator must be purged before use to ensure the patient does not receive an excess level of NO • Loss of communication between the INOmax DS the INOMAX cylinder for more than one hour will result in interruption of INOMAX delivery.
Page 53
Changing INOMAX Cylinders B. Perform high pressure leak test. C. Purge the high pressure hose. D. Connect the pressure hose. Part No. 20751 Rev-01 2014-08...
Page 54
Changing INOMAX Cylinders Changing INOMAX Cylinders (cont.) E. Open the cylinder valve (this may activate the “Two Cylinders Open” alarm until the empty cylinder valve is closed). Note: If using the INOmax DS Transport Regulator/Cap Assembly, transfer the cap from the exhausted INOMAX cylinder to the new INOMAX cylinder at this time;...
Page 55
Changing INOMAX Cylinders (Intentionally left blank) Part No. 20751 Rev-01 2014-08...
High Calibration Connection Diagrams Connection Diagram for NO and NO High Range Calibration INOmax DS 6a. If the pressure is in the red or black 1. Cylinder Cap zone (0-25 psig) select another INOcal 2. Regulator Seal cylinder. 3. Regulator 7.
Page 57
High Calibration Connection Diagrams Calibration Setup for O High Range Calibration 1. 100% O Source 2. O Tubing 3. 15M x 4.5mm I.D. Adapter 4. Gas Sample Tee 5. Patient Gas Sample Line with Nafion Part No. 20751 Rev-01 2014-08...
INOmax DS Disposable Adapters INOmax DS Patient Circuit Disposables (Note: Graphics not actual size) Adapter, Adapter, 22F X 15M 15M Fits 4.5mm ID Tubing Adapter, 22M/15F X 22M/15F Adapter, Cuff, 22mm ID X 22mm ID Adapter, Gas Sample Tee Adapter, 90 degree Sample Port Bunnell Life Pulse Disposable Disk Filter, 0.5 micron Adapters Convenience Pack...