Mallinckrodt INOMAX DSir Plus Operation Manual page 134

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

Figure 5-5 Menu Screen
Figure 5-6 Alarm History Screen
5-4
A complete list of all alarms that have occured
since the INOmax DS
has been turned ON can
IR
be viewed by pressing the ALARM HISTORY
button on the menu screen (see figure 5-5).
Priority
A yellow dot signifies a low priority alarm, and a
red dot signifies a high priority alarm.
Press the EXIT button or the Return to Previous
Menu button to return to the menu screen.
The duration of an alarm is truncated to the second
or minute "label" displayed in the format "xx label,"
where "xx" and "label" are as defined in the table
below.
Alarm Duration Condition
xx < 1 second
1 second ≤ x < 1 minute
1 minute ≤ x < 61 minutes
xx ≥ 61 minutes
in progress
For example, an alarm that occurred for 38.9
seconds would display as "39 sec," while an
alarm that occurred for 75 minutes would display
as "> 60 min."
Displayed
Label
Value
<1
sec
xx
sec
xx
min
> 60
min
Active
N/A
Part No. 20717 Rev-01
2014-07