Mallinckrodt INOMAX DSir Plus Operation Manual page 140

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

Alarm
8. Delivery Failure
9. Delivery Stopped
5-10
High Priority Alarms
Possible Cause
a. Over-delivery of INOMAX.
- (CALCULATED dose is
>200% of set dose) AND
(CALCULATED dose is
> set dose + 10 PPM) for
12 consecutive seconds,
or
- CALCULATED dose >
100 for 12 seconds
b. The internal NO flow sensor
temperature or measured
flow rate is out of range.
Other internal error
detected (a temporary
internal communication
error that has been
detected).
a. MONITORED NO > 100
ppm for at least 12 seconds
(once MONITORED NO
drops below 100 ppm for 12
seconds, alarm will resolve
and INOMAX delivery will
restart).
b. Drug past expiry date
or drug concentration
mismatch.
c. INOmeter may have failed.
d. INOMAX cylinder valve is
closed.
e. INOMAX cylinder is not
detected.
Recommended Action
INOMAX delivery has been interrupted.
Manually ventilate patient with the INOblender
or
turn integrated pneumatic backup delivery ON
(verify at least five L/min ventilator gas flow in
patient circuit).
If Delivery Failure Alarm Occurs:
1. Use one of the back-up INOMAX delivery
options
2. Turn INOmax DS
to standby, then restart.
IR
3. Once device is ready to restart therapy:
4. Turn integrated back up off (if used) and then
set the INOMAX dose.
5. Confirm delivery and check alarms.
Contact Technical Support.
INOMAX delivery has been interrupted.
Manually ventilate patient with the INOblender
or
turn integrated pneumatic backup delivery ON
(verify at least five L/min ventilator gas flow in
patient circuit).
Replace INOMAX cylinder.
Replace INOMAX cylinder.
Open INOMAX cylinder valve.
Replace INOMAX cylinder.
Contact Technical Support.
Part No. 20717 Rev-01
2014-07