Connecting Inomax Ds Ir Sample Tee To The Bunnell Life Pulse Circuit; Ir Injector Module To The Bunnell Life Pulse Circuit - Mallinckrodt INOMAX DSir Plus Operation Manual

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

Connecting INOmax DS
1. From Patient Box
2. Cut Green tube at midpoint
(approximately six inches from the Life Port Adapter)
3. Patient Gas Sample Line with Nafion
4. Insert Sample Tee
5. Life Port Adapter
6. Endotracheal Tube
Figure 3-25
Connecting INOmax DS
3
4
1. Injector Module Electrical Cable
2. NO/N
Injector Tube
2
3. Gas Out Tube from Vent
4. 15M X 4.5 mm I.D. Adapter
5. 22M/15F X 22M/15F Adapter
Figure 3-26
Part No. 20717 Rev-01
2014-07
Sample Tee to the Bunnell Life Pulse Circuit
IR
1
2
4
3
Injector Module to the Bunnell Life Pulse Circuit
IR
1
5
6
6. Injector Module
7. 15M X 4.5 mm I.D. Adapter
8. Three cm Piece of Green Gas Out Tube
9. One-Way Valve
10. Green Gas Out Tube to Humidifier
5
6
2
7
8
9
10
3-33