Alarm Help

• If an alarm occurs, safeguard the patient first before troubleshooting or repair
WARNING:
procedures.
• Use caution when troubleshooting the INOmax DS
When possible, replace the unit in question, and perform troubleshooting
procedure once the unit is removed from the patient.
If the system fails to operate properly:
1. Check the patient condition and take appropriate action.
2. Use the INOblender (see INOblender Operation Manual) or backup if necessary (see page 3-7).
3. Verify that the system is set up as detailed in 3/ Patient Application.
4. Find a symptom or alarm condition in the troubleshooting table which best describes the problem and
follow the recommended actions to resolve the problem.
If the problem can't be corrected:
Contact Customer Care (1-877-566-9466).
If the INOmax DS
must be returned for servicing:
IR
1. Disconnect the following from the unit (refer to Figures 1-1 and 1-2).
Remove and return with unit:
• Injector module
• Injector module cable
Remove and discard:
• Patient gas sample line
• NO/N
injector tube
2
2. Pack the INOmax DS
loaner packaging materials to protect the system during transit. If the proper packaging is not available,
please contact Customer Care (1-877-566-9466).
3. Make sure the outside of the box is labeled: "FRAGILE MEDICAL EQUIPMENT, HANDLE WITH CARE."
4. Send the unit to the location specified by Customer Care.
Part No. 20717 Rev-01
2014-07
and the accessories as requested by Customer Care. Use the original or service
IR
while in use for a patient.
IR
To activate on-screen alarm help, press the
Alarm Help Button next to the Alarm Silence
button.
5-5