Mallinckrodt INOMAX DSir Plus Operation Manual page 24

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

Setting and making changes on the INOmax DS
Dose settings
Displayed dose settings are 1, 5, 10, 20, 40, 60 and
80 ppm. Each click on the control knob corresponds
to a known change in dose. The incremental dose
per click corresponds to a value dependent upon
the dose range in which the change is made, as
illustrated in the table at right.
Dose Settings Adjustments
1-12
IR
Dose Setting Range Dose Change Per Click
< 1 ppm
1 to 40 ppm
40 to 80 ppm
1. SELECT
(press) a button on the touch
screen associated with the
desired function. (An audible
beep will sound when a button
is selected, and the button will
be displayed in inverse video.)
2. ROTATE
the control wheel clockwise or
counterclockwise to adjust the
value.
0.1 ppm
1 ppm
2 ppm
Part No. 20717 Rev-01
2014-07