Inoblender Test - Mallinckrodt INOmax DSIR Plus MRI Operation Manual

Table of Contents

Advertisement

INOblender Test

3
40 ppm
10
L/min
5
1
6
2
• If the INOmax DS
WARNING:
minutes, depressurize the regulator supply line (see next page "Depressurizing
the Regulator Supply Line").
• If the INOmax DS
minutes, repeat automated or manual purge procedure.
• If the INOmax DS
repeat pre-use procedure.
The INOmax DS
IR
Proceed to Section 3/ Patient Application.
2-10
4
5
Plus MRI is not going to be used on a patient within 10
IR
Plus MRI is not used and is pressurized for more than 10
IR
Plus MRI is depressurized and not used within 12 hours,
IR
Plus MRI is now ready to connect to the patient.
1. Remove oxygen tubing from O
and connect to front of INOblender.
2. Remove the MR injector module from
the pre-use set-up and reconnect the
adapters.
3. On the INOblender, set the INOMAX dose
to 40 ppm and O
flow to 10 L/min.
2
4. Allow monitored values to stabilize (may
take up to 3 minutes) and verify the NO
value on the INOmax DS
the table below:
5. Turn the dose and oxygen flow to zero.
6. Remove the pre-use set-up from the
INOblender.
Acceptable
4
NO Value
Pre-use checkout complete.
flowmeter
2
Plus MRI using
IR
32-48 ppm
Part No. 20568 Rev-02
2015-08