Transport Regulator/Cap Assembly Application - Mallinckrodt INOMAX DSir Plus Operation Manual

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

Transport Regulator/Cap Assembly Application

Before leaving the bedside (intrahospital transport) or hospital (interhospital transport), check
Note:
the INOMAX cylinder for the correct product identity labels, cylinder concentration and expiration
date. Verify cylinder has at least 500 psig (replace if 200 psig or less), and tighten the fitting to the
INOMAX cylinder.
Figure 4-3
Part No. 20717 Rev-01
2014-07
1. Connect a high pressure regulator to an
INOMAX cylinder and hand-tighten the fitting
to the INOMAX cylinder (see Figure 4-3).
Ensure the white plastic tip is in place on
Note:
the regulator connector and not chipped
or cracked. Remove and replace as
necessary (see Replacing the CGA 626 tip
on the INOMAX regulator, page 7-8).
2. Connect the INOMAX regulator hose to
one of the INOMAX inlets on the back of the
INOmax DS
(see Figure 4-3).
IR
4-13