Initial connections

1a
2a
2b
2-2
1b
2b
2a
WARNING:
Note:
2c
1. Confirm the water trap bottle and water
separator cartridge are in place
Connect the patient gas sample line with
Nafion to the sample line inlet port on the
front of the INOmax DS
Check cables and hoses for signs of wear and
damage.
2. Connect one end of the injector module
electrical cable to the injector module and
the other end into the front of the
INOmax DS
2a
.
IR
• Line up the red dot on both the connector
and the injector module before inserting
the connector (see inset detail).
Connect one end of the INOMAX injector
tube to the injector module and the other end
into the front panel of the INOmax DS
Be certain all cables and
hoses are positioned to
help prevent damaging or
occluding them.
• It is recommended to disinfect or
sterilize the injector module prior to
initial setup.
• To remove this type of connector, the
knurled sleeve
2c
must be pulled outward before
removing the connector from the
injector module or the front panel.
.
1a
.
1b
IR
.
2b
IR
on the connector
Part No. 20717 Rev-01