Inoblender Test - Mallinckrodt INOMAX DSir Plus Operation Manual

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

INOblender Test

2
3
40 ppm
5
1
Acceptable
NO Value
• If the INOmax DS
WARNING:
depressurize the regulator supply line.
• If the INOmax DS
repeat purge procedure.
Pre-use checkout complete.
The INOmax DS
is now ready to connect to the patient.
IR
Proceed to Section 3/Patient Application.
9-8
4
3
10 L/min
5
2
32-48 ppm
is not going to be used on a patient within 10 minutes,
IR
is not used and is pressurized for more than 10 minutes,
IR
• Ensure INOblender inlet hose
Note:
is connected to the back of the
INOmax DS
and the Quick-Connect
IR
cover is in place.
• Ensure oxygen supply hose is
connected to O
INOblender.
1. Remove the Pre-Use set-up oxygen tubing
from the oxygen flowmeter and connect it to
the front of the INOblender.
2. Remove the Injector Module from the Pre-Use
set-up and reconnect the adapters.
3. On the INOblender, set the INOMAX dose to
40 ppm and O
flow to 10 L/min.
2
4. Verify values on the INOmax DS
5. Turn the dose and flow to zero and remove the
Pre-Use set-up from the INOblender.
inlet fitting on back of
2
.
IR
Part No. 20717 Rev-01
2014-07