Performance Test - Mallinckrodt INOMAX DSir Plus Operation Manual

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

Performance Test

Set Dose
Acceptable O
Value
2
Acceptable NO
Value
2
Acceptable NO Value
2-8
2. Turn the integrated backup INOMAX delivery to
ON (250 mL/min.).
Verify "Backup ON" alarm occurs.
3. Allow monitored values to stabilize (may take
up to 3 minutes).
Verify the NO and NO
following ranges:
4. Turn the backup INOMAX delivery to OFF.
40 ppm
95% ± 3 %
< 1.5 ppm
35-45 ppm
readings are within the
2
NO = 14-26 ppm
NO
1. Using the pre-use set-up connectors, verify that
the O
flowmeter is set to 10 L/min.
2
2. Press NEXT button to automatically set the
INOMAX dose to 40 ppm.
3. Allow values to stabilize. Compare the
INOmax DS
monitor values to the values in
IR
the table.
4. Performance test is complete.
The INOMAX dose is automatically set to zero.
If a monitored value is outside the range
Note:
indicated, see section 5/ Alarm Help.
≤ 1.0 ppm
2
Part No. 20717 Rev-01
2014-07