Integrated Pneumatic Backup INOMAX Delivery Test
1
6
5
7
8
1. O
Flowmeter (Connected to wall/tank)
2
2. Injector Module Electrical Cable
3. NO/N
Injector Tube
2
4. Patient Gas Sample Line with Nafion
5. O
Tubing
2
6. 15M x 4.5 mm Adapter
7. 22M / 15F x 22M / 15F Adapter
8. Injector Module
9. 300 mm of 22 mm Hose
10. Gas Sample Tee
Figure 2-1
Part No. 20717 Rev-01
2014-07
2
3
4
9
10
1. Assemble pre-use set-up connectors and
tubing (press SHOW DIAGRAM button if
needed).
Set the oxygen flowmeter to 10 L/min. (#1 in
Figure 2-1).
2-7