6/ Calibration
INOMAX can be administered during the sensor calibration process. However,
WARNING:
inspired gases are not monitored and gas monitoring alarms are disabled.
During any calibration process, all other alarms remain active while monitoring alarms are disabled.
Note:
Calibration Area
Part No. 20717 Rev-01
2014-07
To access the calibration menu:
Press the menu button on the main screen to
enter the menu screen (second menu level).
The lower part of the screen displays the low
calibration (Low Cal) and high calibration (High
Cal) buttons.
• Select the Low Cal button to start a low
calibration. The date and time the most
recent low calibration occurred is displayed
below the Low Cal button.
• Select the High Cal button to enter the high
calibration screen. The earliest sensor high
calibration due date is displayed above the
High Cal button.
If the date is flashing, it signifies a calibration is
past due.
To return to the main
Note:
screen, press the return to
the previous level button in
the top right of the screen.
Instructions for completing a low and high
calibration are on the following pages.
6-1