Mallinckrodt INOMAX DSir Plus Operation Manual page 43

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

3a
Part No. 20717 Rev-01
2014-07
3. Connect the INOmax DS
emergency-power-backed hospital-grade
outlet. The power cord must always be
connected to an electrical outlet to maintain
a full battery charge.
Verify the main power supply indicator light is
illuminated (green)
3a
. This indicates that the
power cord is plugged into an electrical outlet.
Keep the power cord off of the
Caution:
ground and away from moving
parts.
4. Load two INOMAX drug cylinders onto
cart and check for correct product identity
labels, cylinder concentration (800 ppm) and
expiration date.
For the CGA-type INOMAX regulator
Note:
connector, ensure the white plastic tip
is not chipped or cracked. Remove and
replace as necessary. (see Replacing the
tip on the INOMAX regulator, Page 7-8).
5. Connect an INOMAX regulator to one of the
INOMAX cylinders, and hand tighten the fitting
to the INOMAX cylinder.
Connect the INOMAX regulator hose to one of
the INOMAX inlets (see page 7-9 for additional
instructions).
If using the transport regulator/cap
Caution:
assembly (PN 10022) see Figure
4-9, Section 4/ Transport.
power cord to an
IR
2-3