Mallinckrodt INOMAX DSir Plus Operation Manual page 111

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

8a
9a
9b
IR Cable to the
INOmax DS
Cart
IR
Part No. 20717 Rev-01
2014-07
7
8b
8a
9b
7. Open INOMAX transport cylinder valve.
8a. Immediately remove "88" INOMAX cylinder
regulator hose, and close cylinder valve.
8b. Depressurize the "88" INOMAX cylinder
regulator hose by pressing in the purge
port.
9a. Disconnect the cart infrared (IR) cable.
9b. Connect the IR cable from the transport
regulator cap assembly to the back of the
INOmax DS
.
IR
4-3