• The INOmax
adds NO/N
IR
at 80 ppm) and subtracts gas from the breathing circuit via gas sampling at a nominal flow rate of 0.23 L/min.
• These effects impact the delivered gas flow rate when using the Vapotherm 2000i. It is recommended that after an
NO setting change the user checks the delivered gas flow rate and adjusts the gas source flow rate as necessary.
1. INOmax DS
IR
2. O
Flowmeter
2
3. O
Tubing
2
4. 15M x 4.5 mm Adapter
5. 22M/15F x 22M/15F Adapter
6. 300 mm of 22 mm Hose
7. 22M/15F x 22M/15F Adapter
8. Injector Module
9. 15M x 4.5 mm Adapter
10. Vapotherm 2000i
11. Patient Delivery Tube
12. O
Tubing Sample Tee
2
13. Patient Cannula
14. Patient Gas Sample Line with Nafion
15. NO/N
Injector Tube
2
16. Injector Module Electrical Cable
Figure 3-40 Example: Vapotherm 2000i Circuit Diagram
3-48
gas flow to the breathing circuit flow in proportion to the NO setting (up to 10%
2
16
15
14
13
1
12
11
2
3
4
5
6
7
8
9
10
Part No. 20717 Rev-01