Viasys Infant Flow Sipap - Mallinckrodt INOMAX DSir Plus Operation Manual

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

Viasys Infant Flow SiPAP

• The INOmax DS
adds NO/N
IR
(up to 10% at 80 ppm) and subtracts gas from the breathing circuit via gas sampling at a nominal
flow rate of 0.23 L/min.
• These effects change the flow going to the nasal adapter and can therefore impact the CPAP level
established by specific flow settings (See table below). The maximum flow error is approximately
11% at two L/min which is within the accuracy of the flow meter specification (+/-15%).
• It is recommended that after an NO setting change the user checks the CPAP level on the Infant
Flow SiPAP front panel display and adjusts as necessary.
SiPAP Flow Setting
DS
After INOmax
Set @ 0 ppm
IR
% error
DS
Set @ 80 ppm 1.97
After INOmax
IR
% error
3-52
gas flow to the breathing circuit flow in proportion to the NO setting
2
Flow
Flow
(L/min)
(L/min)
2
4
1.77
3.77
-11.5%
-5.8%
4.17
-1.5%
4.3%
Flow
Flow
Flow
(L/min)
(L/min)
(L/min)
6
8
10
5.77
7.77
9.77
-3.8%
-2.9%
-2.3%
6.37
8.57
10.77
6.2%
7.1%
7.7%
Part No. 20717 Rev-01
2014-07