7/ Maintenance
• To help prevent fire, use only lubricants approved for O
Caution:
• Do not use lubricants which contain oil or grease, as they burn or explode in high O
concentrations.
• Do not sterilize or disinfect with the power connected.
Note:
The INOmax DS
User Maintenance Schedule
Frequency
Maintenance
Daily
1. Check the INOMAX cylinder pressure. A cylinder with less than 200 psig should
be replaced.
2. Empty the water trap bottle as needed.
Start of each
Must perform the Pre-Use Procedure.
patient
Between each
1. Sterilize and/or disinfect the Injector Module.
patient
2. Clean water trap bottle.
3. Replace the single patient-use items.
4. Make sure that the delivery system power cord is always plugged into an
5. Make sure the connectors, hoses and cables are in good condition.
Monthly
1. Perform a low and a high calibration of NO, NO
Note:
2. Check INOMAX regulators for leaks.
Part No. 20717 Rev-01
2014-07
does not contain any user repairable parts.
IR
emergency-power-backed electrical outlet.
A flashing date above the high sensor calibration button signifies a high
calibration is past due.
equipment, such as KRYTOX
2
and O
.
2
2
.
®
2
7-1