Intrahospital Transport (Within The Hospital) When Removing The Inomax Ds Ir And Inoblender From The Cart - Mallinckrodt INOMAX DSir Plus Operation Manual

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

B. Intrahospital transport (within the hospital) when removing the
INOmax DS
and INOblender from the cart.
IR
4
4-2
3
2
1
2
4
5
6a
6b
Check the INOMAX drug cylinders for the
correct product identity, cylinder concentration,
and expiration date.
1. Connect the high pressure transport
regulator/cap assembly to the INOMAX
transport cylinder, and tighten the fitting to
the INOMAX cylinder.
2. Place the cap assembly over the INOmeter.
3. Open then close the INOMAX transport
cylinder valve.
Be sure to align the key way inside
Note:
the Cap Assembly with the ibutton on
the INOmeter.
4. Check for adequate cylinder pressure. Verify
cylinder has at least 500 psig (replace if 200
psig or less).
5. Monitor pressure gauge for 30 seconds for
any signs of pressure decrease. If no signs
of pressure decrease is observed, the leak
test is successful.
6a. Depressurize the transport
cylinder regulator hose.
6b. Connect regulator hose to
the available drug inlet.
Part No. 20717 Rev-01
2014-07