Mallinckrodt INOMAX DSir Plus Operation Manual page 53

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

3/ Patient Application
Before Operation
Complete the initial connections and Pre-Use Checkout procedure as described in the previous sections
before connecting the INOmax DS
manual for its setup and operation)
• Set the INOmax DS
WARNING:
monitor any inadvertent changes in treatment. For alarm information,
see Section 5/ Alarms.
• The use of pediatric and neonatal ventilator settings with adult size breathing
circuits can result in high levels of NO
that is appropriate for the patient.
• Abrupt discontinuation of INOMAX may lead to worsening oxygenation and
increasing pulmonary artery pressure, i.e., Rebound Pulmonary Hypertension
Syndrome. To avoid abrupt discontinuation, utilize the INOblender or integrated
pneumatic backup if necessary. If Rebound Pulmonary Hypertension occurs,
reinstate INOMAX therapy immediately (See the INOMAX prescribing Information
for further details.)
• If the high NO
proper set up while maintaining INOMAX delivery. Adjust INOMAX and/or FiO
as appropriate. (See INOMAX Prescribing Information for further details on
the effects of Nitrogen Dioxide, NO
increased NO
Use Outside of Product Labeling
• The INOmax DS
contraindications, warnings and precautions described in the INOMAX (nitric
oxide) drug package inserts and labeling. Refer to this material prior to use.
• The manufacturer does not recommend that the INOmax DS
helium/oxygen mixtures in any situation. The INOmax DS
INOMAX therapy gas only in conjunction with the delivery of air and oxygen. The
use of helium/oxygen mixtures will lead to over-delivery of INOMAX which may
lead to interruption of therapy.
• The use of devices which radiate high-intensity electrical fields may affect the
operation of the INOmax DS
support equipment is mandatory whenever interfering devices are in operation
on or near a patient.
• Do not connect items which are not specified as part of the system.
• The approved patient population for the INOmax DS
labeling for INOMAX (nitric oxide) for inhalation, is limited to neonates. The
INOmax DS
• Use distilled water in the humidifier to prevent the formation of bases or acids.
Caution:
• High frequency and/or high intensity light emission, in the area of the INOmeter, may
interfere with communication between the INOmax DS
cylinder (see page 1-16).
• The gas sensors in the INOmax DS
gas to function correctly over the long term. Using the INOmax DS
humidifier will shorten the life of the gas sensors.
Part No. 20717 Rev-01
2014-07
into the patient's breathing circuit. (See the ventilator/breathing device
IR
alarm thresholds for the current patient conditions to
IR
alarm activates, the delivery system should be assessed for
2
levels, call technical support, do not discontinue therapy.
2
must only be used in accordance with the indications, usage,
IR
. Constant surveillance of all monitoring and life
IR
is not intended to be used in other patient populations.
IR
monitoring system require humidity in the sample
IR
. Always use the size of breathing circuit
2
). If unable to determine the cause of the
2
IR
, as specified in the drug
IR
and the INOmeter on the INOMAX
IR
IR
2
be utilized with
IR
is intended to deliver
without water in the
3-1

Advertisement

Table of Contents
loading

Table of Contents