Mallinckrodt INOMAX DSir Plus Operation Manual page 115

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

5
8
Part No. 20717 Rev-01
2014-07
4
6
7
The INOblender is now ready for patient use.
4. Verify INOMAX cylinder valve is open.
5. Verify cylinder has at least 500 psig (replace if
200 psig or less)."
Adjust Settings
6. Turn the INOblender setting dial to the
desired concentration (5 to 80 ppm for an
800 ppm cylinder).
7. Turn the O
flowmeter to the desired flow rate
2
(5 to 14 L/min).
8. Squeeze the manual resuscitator 3-4 times
to purge the NO
from the system.
2
4-7