Mallinckrodt INOMAX DSir Plus Operation Manual page 153

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

Part No. 20717 Rev-01
2014-07
3. When the low calibration is successful, a single
tone will be heard and the main screen will
appear. A two minute Monitor Alarm Delay
Active indicator will occur, preventing monitoring
alarms from occurring while the measured value
stabilizes. All system alarms are still active.
If the low calibration was unsuccessful, the
INOmax DS
will automatically attempt another
IR
low calibration. If the second low calibration attempt
fails, the alarm area will display a Low Calibration
Failed alarm.
• Attempt a manual low calibration.
If a sensor has failed, the display will indicate the
failed sensor symbol in the monitoring area of
that sensor. (Press the Alarm Help
button for on-screen alarm help or see
Section 5/ Alarms.)
6-3

Advertisement

Table of Contents
loading

Table of Contents