Mallinckrodt INOMAX DSir Plus Operation Manual page 199

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

3
Part No. 20717 Rev-01
2014-07
3. Connect the INOmax DS
emergency-power-backed hospital-grade
outlet. The power cord must always be
connected to an electrical outlet to maintain
a full battery charge.
Verify the main power supply indicator light
is illuminated (green). This indicates that the
power cord is plugged into an electrical outlet.
Keep the power cord off of the
Caution:
ground and away from moving
parts.
4. Load two INOMAX drug cylinders onto
cart and check for correct product identity
labels, cylinder concentration (800 ppm) and
expiration date.
5. Connect an INOMAX regulator to one of the
INOMAX cylinders, and hand tighten the fitting
to the INOMAX cylinder. Connect the INOMAX
regulator hose to one of the INOMAX inlets
(see page 7-9 for additional instructions).
If using the transport regulator/cap
Caution:
assembly (PN 10022) see Figure
4-7, Section 4/ Transport.
For the CGA-type INOMAX regulator
Note:
connector, ensure the white plastic tip
is not chipped or cracked. Remove and
replace as necessary. (see Replacing the
tip on the INOMAX regulator, Page 7-8).
power cord to an
IR
9-3

Advertisement

Table of Contents
loading

Table of Contents