adds NO/N
IR
to 10% at 80 ppm) and subtracts gas from the breathing circuit via gas sampling at a nominal flow rate
of 0.23 L/min.
These effects impact the delivered gas flow rate when using the Vapotherm Precision Flow.
It is recommended that after an NO setting change the user checks the delivered gas flow rate and
adjusts the gas source flow rate as necessary.
Follow all manufacturer instructions for connection to the Vapotherm Precision Flow.
1
1. Patient Gas Sample Line with Nafion
2. INOmax DS
3. Precision Flow Unit
4. Injector Module
5. Patient Delivery Tube
Figure 3-41 Example: Vapotherm Precision Flow Circuit Diagram
Part No. 20717 Rev-01
2014-07
gas flow to the breathing circuit flow in proportion to the NO setting (up
2
2
8
9
7
6
IR
3
4
5
6. Oxygen Tubing Sample Tee
7. Patient Cannula
8. Injector Module Electrical Cable
9. NO/N
Injector Tube