Mallinckrodt INOMAX DSir Plus Operation Manual page 113

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

Once the devices are secure on the cart:
17a. Open/close "88" INOMAX cylinder, then depressurize the regulator hose by pressing the
regulator hose into the purge port.
17b. Open the "88" INOMAX cylinder valve and insert the regulator hose into available INOMAX gas
inlet.
17c. Close transport cylinder valve and remove the regulator hose from the INOMAX gas inlet and
depressurize the regulator hose by pressing into the purge port.
17d. Disconnect the transport cylinder IR cable from the back of the INOmax DS
17e. Attach IR cable from the cart to the back of the INOmax DS
17a
17b
Part No. 20717 Rev-01
2014-07
17a
17c
17b
17e
.
IR
.
IR
17d
17c
4-5