Mallinckrodt INOMAX DSir Plus Operation Manual page 194

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

portable and mobile RF communications equipment and the
The
INOmax DS
system
IR
disturbances are controlled. The user of the
interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the
output power of the communications equipment.
Rated
Maximum
Output
150 kHz to 80 MHz
Power of
Outside ISM bands
Transmitter
d=1.2*
W
0.01
0.12
0.1
0.38
1
1.2
10
3.8
100
12
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter,
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
NOTE 3: An additional factor of 10/3 is used in calculating the recommended separation distance for
transmitters in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency
range 80 MHz to 2.5 GHz to decrease the likelihood that mobile/portable communications
equipment could cause interference if it is inadvertently brought into patient areas.
NOTE 4: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
8-12
Recommended separation distances between
is intended for use in an electromagnetic environment in which radiated RF
INOmax DS
INOmax DS
system
IR
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
In ISM bands
P
d=1.2*
P
0.12
0.38
1.2
3.8
12
INOmax DS
system
can help prevent electromagnetic
IR
as recommended below, according to the maximum
m
80 MHz to 800 MHz
d=1.2*
P
0.12
0.38
1.2
3.8
12
system
IR
800 MHz to 2.5 GHz
d=2.3*
P
0.23
0.73
2.3
7.3
23
Part No. 20717 Rev-01