Dräger Babylog VN500/Infinity Acute Care System and Heinen & Löwenstein
Leoni-plus Ventilator
WARNING:
Omission of the one-way valve may result in high NO delivery.
• Validated for use outside of the United States.
Note:
• Use a one-way valve (Part Number- 1605-3139-000) when ventilating during the HFOV mode.
1
15
1. Patient wye
2. Dräger Babylog VN500 / Leoni-plus Ventilator
3. Ventilator Expiratory Port
4. Ventilator Inspiratory Port
5. Patient Gas Sample Line Input Connection
6. INOmax DS
IR
7. NO/N
Injector Tube Front Panel Connection
2
8. Injector Module Electrical Cable Front Panel Connection
Figure 3-28 Example: Dräger Babylog VN500 and Leoni-plus Circuit Diagram
3-36
2
3
4
14
13
12
5
11
10
9
9. Injector Module
10. One-Way Valve
11. Humidifier Inlet
12. Humidifier
13. Humidifier Outlet
14. Patient Gas Sample Line with Nafion
15. Gas Sample Tee
6
7
8
Part No. 20717 Rev-01