No Sensor High Calibration - Mallinckrodt INOMAX DSir Plus Operation Manual

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

NO Sensor High Calibration

• When performing a high calibration, make sure to select the correct calibration gas and
Caution:
confirm the expiration date before using.
• Never connect the sample line directly to a high pressure gas source (greater than
150 cmH
Note:
Complete a low calibration (see Low Calibration section) prior to completing the high calibration.
One-way Valve
Gas Flow
4
1
Figure 6-1
Part No. 20717 Rev-01
2014-07
O); this could damage the sampling system.
2
5
2
3
INOcal calibration gas kit sample tubing
When using the calibration tubing kit
(P/N 50239), which is supplied with the INOcal
regulator kit (P/N 10090), ensure that the beige
pressure relief valve supplied with the tubing is
installed and oriented as indicated in Figure 6-1.
An incorrectly installed one-way
Caution:
valve can lead to over-pressurization
of the sampling system. A leak in
the calibration tubing kit (PN 50239)
attached to the calibration cylinder
regulator can result in displayed NO
values greater than the set dose
value after passing a low and high
calibration successfully. This can be
caused by aging of the calibration
tubing. The calibration tubing kit
should be replaced under the
following circumstances:
• The tubing is discolored or stiff.
• There is a crack or break in the
tubing.
1. Remove cylinder cap and inspect for damaged
threads and contaminants.
2. Check seal on regulator. Verify it is correctly in
place and undamaged.
3. Attach regulator to cylinder (Turn regulator nut
counter-clockwise to tighten).
4. NOTE: Tubing adapter will be required if
regulator outlet diameter measures 0.24 in.
(6.10 mm).
5. Attach tubing kit to regulator outlet.
NOTE: Confirm water trap bottle, water separator
cartridge and patient gas sample line are in place.
6-7