Mallinckrodt INOMAX DSir Plus Operation Manual page 147

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

Indicator
32. NO Delivery Button Inactive
33. High Calibration Cancelled a. If a high calibration is
34. High Calibration Failed
Part No. 20717 Rev-01
2014-07
Indicators
Possible Cause
a. Device does not recognize
an INOMAX cylinder, the
dose knob will be greyed
out and it will not allow
the user to set an initial
INOMAX dose.
b. IR cable is not connected
to the back of the
INOmax DS
.
IR
c. Integrated pneumatic
backup switch is ON.
cancelled the progress bar
will turn yellow.
a. If a high calibration fails, the
progress bar will turn red.
Recommended Action
1. Load INOMAX cylinder on to the cart.
2. Remove any obstruction between the
INOmeter and the INOmax DS
3. Move the interfering light source.
4. Move the INOmax DS
IR
the high intensity light in the area of the
INOmeter.
5. Shield the INOmeter from the suspect light
source.
Verify the IR cable is connected to the back of
the INOmax DS
.
IR
1. Correct the reason for initiating integrated
pneumatic backup delivery.
2. Turn integrated pneumatic backup delivery
OFF and confirm set NO dose has been
restored.
Restart high calibration.
or
Exit calibration screen and return to main
screen.
Repeat the high calibration.
Contact Technical Support.
cart cover.
IR
cart to reduce
5-17