• Confirm INOblender inlet hose is
Note:
connected to the INOMAX regulator
hose and the Quick-Connect cover
is in place.
• Confirm 50 psig oxygen supply
hose is connected to O
on back of INOblender.
1. Use the Pre-Use set-up to connect to
sample gas from the INOblender.
Connect O
tubing to the front of the
2
INOblender.
2.Confirm the injector module is removed
from the Pre-Use set-up and connect the
adapters.
3. On the INOblender, set the INOMAX dose
to 40 ppm and O
flow to 10 L/min.
2
4. Verify NO value on the INOmax DS
5. Turn the dose and oxygen flow to zero.
6. Remove the Pre-Use set-up from the
INOblender.
inlet fitting
2
.
IR
4-9