INOblender Circuit
Connection Diagram
INOblender Warnings:
• The purge procedure must be followed to help ensure NO
before the manual resuscitator bag is connected to the patient.
The manual bag should be squeezed continuously during use
to avoid NO
building up in the bag. If the bag is not squeezed
2
removed from the patient and the purge procedure performed
before continuing.
• Persons using this device should be trained on and
experienced in the use of this device to assure effective
administration of INOMAX and to avoid injury to the patient or
dioxide or other reaction products.
INOblender Cautions:
• Refer to the manufacturer's procedures for using the resuscitation
until the NO pressure gauge reads zero, then turn the O
2
Note: Connections to various ventilators as well as their corresponding
disposable circuits, are unique to each manufacturer. Please
instructions for use for guidance.
Part No. 20751 Rev-01
2014-08
50