Mallinckrodt INOMAX DSir Plus Operation Manual page 162

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

6a
6
9c
9b
6-12
7
To INOmax DS
IR
9a
6. Open the INOcal cylinder valve (turn counter-
clockwise) to flow gas.
6a. If the pressure is in the red or black zone (0-
25 psig) select another INOcal cylinder.
7. Attach tubing kit to patient gas sample line.
Cap off sample tee if patient gas sample
Note:
line is removed from patient breathing
circuit.
If the date is flashing on the high
Note:
calibration screen, the calibration is past
due.
8. From the high calibration screen
(third menu level), press the 10 ppm
NO
button to initiate the NO
2
calibration.
9a. To continue the high calibration wizard, press
the NEXT button.
9b. To start the high calibration without using the
wizard, press the START CAL button.
9c. To exit the NO
high calibration wizard, press
2
the CANCEL button.
high
2
Part No. 20717 Rev-01