Replacing The O - Mallinckrodt INOMAX DSir Plus Operation Manual

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

Replacing the O

Figure 7-4
Figure 7-5
NO
Shorting Wire
2
Part No. 20717 Rev-01
2014-07
, NO and NO
2
Sensors
(cont'd)
2
Replace the sensor cover and tighten the two
screws clockwise (see Figure 7-4).
Note:
Newly Installed Sensor
O
and NO
2
2
NO
Insufficient conditioning will result in inaccurate
gas readings.
Perform a low and high calibration for the sensor
before returning the system to use.
Time to Condition Prior
to Calibration
40 minutes
5 hours
7-7