Mallinckrodt INOMAX DSir Plus Operation Manual page 155

800 ppm inomax (nitric oxide) for inhalation series 3 software
Table of Contents

Advertisement

Part No. 20717 Rev-01
2014-07
4. After reaching step 5 of the calibration wizard,
the calibration will take approximately three
minutes, during which a progress bar for the O
sensor indicates progress.
If the CANCEL button is pressed during
Note:
the high O
calibration process, the
2
calibration will be cancelled and the user
will be returned to the high calibration
screen.
5. When the O
high calibration is successful, a
2
single tone will be heard and the progress bar will
turn green. Press the NEXT button to return to
the high calibration screen.
The monitor displays should indicate
Note:
approximately 100% O
and 0.0 ppm NO.
Disconnect the sample line from the calibration
setup and turn OFF the O
Reconnect patient gas sample line
Note:
if previously removed from patient
breathing circuit.
A Monitor Alarm Delay Active indicator will occur
anytime the user exits the calibration screen, even
if a calibration was not initiated. All other alarms
are still active (see section 5/ Alarms for additional
information).
When using the high calibration wizard, if the
BACK button is pressed during the O
calibration process, the progress bar will turn
yellow, signifying a cancelled calibration.
• Press the NEXT button to start the high
calibration wizard.
• Press the START CAL button to start the
calibration again.
• Press the CANCEL button to exit the O
calibration screen.
2
, 0.0 ppm NO
2
2
flowmeter.
2
high
2
high
2
6-5