9-6
Triggering
Table 9-1
Auto delay settings
Volts
Range
2V
20V
200V
Measure action
The measure action block of the trigger model is where a measurement is performed. How-
ever, if the repeating filter is enabled (see Figure 9-3), the instrument samples the specified num-
ber of reading conversions to yield single filtered reading. Only one reading conversion is
performed if the digital filter is disabled, or after the specified number of reading conversions
for a moving average filter is reached.
Figure 9-3
Measure action block
of trigger model
Output triggers
Model 6514 can send out an output trigger (via the rear panel TRIGGER LINK connector)
right after the measure action and/or when operation leaves the trigger layer. An output trigger
can be used to trigger another instrument to perform an operation (e.g., select the next output
step for a source).
Amps
Delay
Range
Delay
5ms
20pA
2500ms
3ms
200pA
2500ms
2ms
2nA
10ms
20nA
10ms
200nA
10ms
2µA
10ms
20µA
5ms
200µA
5ms
2mA
1ms
20mA
0.5ms
Measure Action
Filter Process (Repeat)
CONV
CONV = Reading Conversion
Ohms
Range
Delay
2kΩ
5ms
20kΩ
1ms
200kΩ
1ms
2MΩ
10ms
20MΩ
10ms
200MΩ
10ms
2GΩ
50ms
20GΩ
50ms
200GΩ
50ms
CONV
CONV
Coulombs
Range
Delay
20nC
3ms
200nC
3ms
2µC
3ms
20µC
3ms