Table 3-2
DC output commands
Command
CLEar
CURRent:RANGe <n>
CURRent:RANGe:AUTO <b>
CURRent <n>
CURRent:COMPliance <NRf>
CURRent:FILTer <b>
OUTPut:RESPonse <name>
OUTPut <b>
SourceMeter
1. To select a fixed source range, specify the current output value that is going to be sourced. The
Model 622x will go to the lowest range that can source that value. For example, If you are going
to source 25mA, let <n> = 25e-3. The 100mA range will be selected.
2. Selecting a fixed source range disables autorange.
3. The output must be off in order change the output response. Sending this command while the
output is on will generate error -220 Execution Error.
4. OUTP OFF turns the output off, but does change the set output level. SOUR:CLE sets the out
put level to zero and then turns the output off.

Programming example

The following programming example shows a typical command sequence to con-
figure and control the DC output:
CLEar
CURRent:RANGe:AUTO ON
CURRent 12e-3
CURRent:COMPliance 10
OUTPut ON
OUTPut OFF
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Description
Turns output off and sets output level to zero.
Sets current source range (amps).
<n> = -105e-3 to 105e-3
Enables or disables source autorange.
<b> = ON or OFF
Sets DC current source output level (amps).
<n> = -105e-3 to 105e-3
Sets voltage compliance (volts).
<NRf> = 0.1 to 105
Enables or disables the output analog filter.
<b> = ON or OFF
Select fast or slow output response speed for 6221.
<name> = FAST or SLOW
Turn output on or off.
<b> = ON or OFF
Sets output to zero, then turns the output off.
' Turns the output off.
' Enables autorange.
' Sets output level to +12mA.
' Sets voltage compliance to 10V.
' Turns the output on.
' Turns the output off.
Return to
Section 3 topics
DC Current Source Operation 3-17
1, 2
3, 4
4
Default
100e-3
OFF
0.0
10.0
OFF
FAST
OFF