Troubleshooting the Installed Instrument

Troubleshooting the Installed Instrument
Here is a list of some easily corrected troubleshooting cases. If the fault is not described
below, investigate further without delay. For more information, see "Testing the Instrument"
and "Repairing the Instrument".
Troubleshooting with Self-Test Alarm Messages
(When You Switch the Monitor On)
Message
Battery Low (approxi-
mately 20 minutes remain-
ing)
Check Status Log
ECG EQUIP MALF
NBP EQUIP MALF
Pressure Zero & Check
Cal
PRESS EQUIP MALF
RESP EQUIP MALF
210
What To Do
Connect to AC power to charge the battery, or fit a
fully charged battery within next 15 minutes.
This indicates a non-critical problem in the
Monitor. There is a defect but the Monitor can still
be used.
The Monitor should be investigated technically,
however, as soon as possible.
Measurement defective. Exchange the
Measurement Server.
The Measurement Server should be investigated
technically as soon as possible.
Measurement defective. Exchange the
Measurement Server.
The Measurement Server should be investigated
technically as soon as possible.
Pressure must be zeroed, or calibration required.
Zero the Pressure, or check the calibration.
If unsuccessful, exchange the Measurement Server.
Measurement defective. Exchange the
Measurement Server or Measurement Server
Extension.
The Measurement Server or Measurement Server
Extension should be investigated technically as
soon as possible.
Measurement defective. Exchange the
Measurement Server.
The Measurement Server should be investigated
technically as soon as possible.