Installing the Measurement Server Mounting Options
Step 6
Step 7
Step 8
Step 9
110
Select two screws and suitable fixings appropriate to surface the material. If
necessary, mark the screw positions on the surface and drill suitable holes.
Insert the 2 screws through the screw holes provided in the plate.
Tighten the screws.
Align the feet on the MMS with the locating shoes (7) on the mounting plate and
slide the MMS onto the mounting plate. The Server snap-fits onto the plate.
7
To release the Server, press and hold the latch (in the middle at the top of the mount)
away from the Server and slide it out.