Installing the 12V Adapter

The 12V adapter (option number M3080A #C32) is used with a vehicle 12V supply to power
an M3/M4 Monitor. The Adapter must be connected to ground or to the vehicle chassis.
Detailed instructions can be found in the Instruction sheet delivered with the Adapter
(M3080-9011A). After installation, a safety test must be performed as described in Chapter 4,
"Testing and Inspecting".

Using the Battery Charger and Conditioner

The Battery Charger and Conditioner (M8043A, part number M8043-60000) is an external
device used to recharge and condition the monitor's battery. To use the Battery Charger and
Conditioner, follow the instruction sheet delivered with it. For more information on battery
use, see the "Using Your Monitor in Patient Transport" chapter in the Instructions for Use.

Connecting a Printer

Selecting a Printer

If you are printing locally, you can use either of the following accessories:
• a PCL 3 or PCL 5 compatible printer, with an infrared-to-parallel converter (JetEye 
Printer wireless infrared connector), which you can order as M3080A Option #H05, or
• a PCL 3 or PCL 5 compatible with a built-in infrared interface.
You can also use the LaserJet with the JetEye, but then you will also need a Centronics
printer extension cable. (Installing the wireless infrared printer connector is covered on
page 94).
Make sure that the infrared printer port is at least 50cm (20 inches) from any SpO
while you are printing, to avoid disturbing the SpO
Note that the M3 Print Server is no longer available, but is mentioned here for those who still
use it.
Step 1
Step 2
Step 3
Step 4
Step 5
Step 6
Installing the Instrument
Press the Setup key.
Move the highlight to Printer.
Press on the strip to select the Printer window.
Move the highlight to Printer.
Press on the strip.
Select the appropriate setting according to the following table:
None
If there is no printer. All printing operations are disabled.