Accuracy and Performance Procedures
Respiration Performance
Step 5
Step 6

Invasive Pressure Performance Test

This test checks the performance of the invasive pressure parameter. The Invasive Pressure
performance test is required once every two years.
Tools required: Patient Simulator.
Step 1
Step 2
Step 3
Step 4
Step 5
Step 6
NOTE
If the Measurement Server was calibrated with a dedicated reusable catheter, check the
calibration together with this catheter.
156
Change the Patient Simulator configuration to:
— Base impedance line 1500 Ohm.
— Delta impedance 0.5 Ohm
— Respiration rate 40 rpm.
The value should be 40 rpm +/- 2 rpm.
Connect the Patient Simulator to the Pressure connector on the Measurement Server
or the Measurement Server Extension.
Set Patient Simulator to 0 pressure.
Make a zero calibration.
Configure the Patient Simulator as follows:
— P(static) = 200 mmHg.
Wait for the display.
The value should be 200 mmHg ± 5 mmHg. If the value is outside these tolerances,
calibrate the Measurement Server or Measurement Server Extension.
Testing the Instrument

Advertisement

Table of Contents

   Also See for Philips M3001A

This manual is also suitable for:

M3000aM3046aM3015aM3016a