Recommendations for Test Frequency

The testing checklist appears in the next section of this chapter. The listed procedures should
be performed as indicated in the Suggested Testing Timetable below. The checklist may be
photocopied and should be completed by the tester. It should be filed for future reference.
Suggested Testing Timetable
Functionality Assurance
Preventive Maintenance Tests
Performance and Safety Tests
including Safety Checks (in
accordance with IEC 60601-1)
Testing the Instrument
Performance Assurance Test
System Check
System Self-Test
NBP Calibration
Sidestream CO
Calibration
2
Temperature Accuracy
ECG/Resp Performance
Invasive Pressure Performance
SpO
Performance
2
Mainstream CO
Performance
2
Nurse Call Relay Performance
ECG Sync Performance
System Enclosure Leakage
Current
Protective Earth
Patient Leakage Current
Recommendations for Test Frequency
Frequency
To verify proper operation when
the monitor has been installed.
an assembly has been replaced.
to verify the operation any time the
unit is suspect.
Once every two.
Once a year, or following any
Instrument repairs or the replacement
of any Instrument parts (applies to
units with serial number prefix
DE020xxxxx only).
Once every two years.
Once every two years.