Philips M3001A Service Manual page 196

M3/m4 monitors,measurement server,measurement server extensions
Table of Contents

Advertisement

Part 2 Isolating and Solving Instrument Problems
CO
(M3016A)
CO
(M3016A)
CO
WARM UP
CO
(M3016A)
CO
NING
(M3016A)
CO
(M3016A)
CO
SCALE
(M3015A)
180
CAL FAILED
2
CHECK CAL
2
SENSOR
2
WAIT CAL 2
2
CAL RUN-
2
CAL MODE
2
CHANGE
2
CO
UPDATE FW
2
Calibration aborted due
to power failure, unsta-
ble signal during cali-
bration or transducer
placed on the wrong
calibration cell.
CO
value is less than
2
-4mmHg, greater than
150 mmHg.
The sensor has not
reached operating tem-
perature.
The CAL 1 calibration
cycle is complete; the
CAL 2 calibration cycle
must be started.
CO
calibration is run-
2
ning.
Cal mode is set but the
calibration has not been
initiated.
ETCO
wave is larger
2
than the display chan-
nel.
The software in the
Measurement Extension
does not match the soft-
ware in the Measure-
ment Server. This is
only likely to occur
after a repair or
upgrade.
1. Verify power source.
2. Perform calibration
procedure again.
3. Replace transducer
and repeat calibration
again.
4. Replace the exten-
sion.
Check that calibration
values are within speci-
fied limits by perform-
ing an accuracy check.
Calibrate, if required.
Allow the sensor to
warm up.
Place sensor on other
cal cell and start the
CAL 2 calibration
cycle.
Wait for CO
calibra-
2
tion to complete.
Initiate the calibration
or switch off cal mode
Increase the scale of the
display channel.
Perform the Firmware
update in CO
Service
2
Mode:
select "Update
Firmware" and follow
the instructions on the
display.
In certain situations, no
update is possible - in
these cases, continued
operation is possible
despite the INOP.