Non-invasive Blood Pressure (NBP) Measurement
NBP Modes
The measurement offers adult, pediatric, and neonatal modes.
The following table lists the cuff inflation limits for each mode:
Mode
Adult
Pediatric
Neonatal
The following table lists the measurement ranges for each mode:
Mode
Adult
Pediatric
Neonatal
Safety
The following table lists the maximum limits that ensure patient safety:
Mode
Adult
Pediatric
Neonatal
If any one of these safety limits is violated, an INOP is generated and the valve opens.
30
Subsequent Inflations,
First Inflation
Above Systolic Pressure
165
25
130
20
100
15
Systolic
Diastolic
30 - 270
10 - 245
30 - 180
10 - 150
30 - 130
10 - 100
Maximum
Maximum Time/
Measurement Time
Pressure
180 seconds
180 seconds for
pressure > 15
mmHg
180 seconds
180 seconds for
pressure > 15
mmHg
90 seconds
90 seconds for pres-
sure > 5 mmHg
Stat Mode
15
15
15
Mean
20 - 255
20 - 160
20 - 120
Over-pressure
Maximum
300 mmHg for > 2
seconds
300 mmHg for > 2
seconds
150 mmHg for > 2
seconds
Introduction to the Instrument