Philips M3001A Service Manual page 239

M3/m4 monitors,measurement server,measurement server extensions
Table of Contents

Advertisement

Step 7
Step 8
Step 9
Step 10 Use a small screwdriver to remove the Monitor Bezel assembly cable:
NOTE
Do not touch the LCD display
.
Repairing the Instrument
Lay the Monitor on its rear panel and pull out the two latches - there is one on each
side of the Monitor Bezel:
Stand the Monitor up again and gently lift the Monitor Bezel up and out. Gently let
it rest in front of the Monitor. Be careful. It still has a cable attached:
Remove the white plastic clamp securing the ribbon cable connections. To remove,
push the latch at the top left corner to the left.
Disassembly for the Monitor
223