Troubleshooting the Installed Instrument
Message
"Measurement Server
Configuration not
supported" status message
"Some measurements
in MMS Extension are
not supported" prompt
message
"Measurement Server
not supported" status
message
212
What To Do
Either
A Measurement Extension (M3015A or
M3016A) is connected to a standard M3000A
Measurement Server (noninvasive measure-
ments only) and an M3046A Monitor with soft-
ware Rev.D or lower. No measurements (for
example CO
, 2nd Press/Temp) from the Meas-
2
urement Extension are available.
Exchange the Measurement Server for an
M3000A with option #C06.
Or
A Measurement Extension (M3015A or M3016A)
is connected to a standard M3001A Measurement
Server (noninvasive measurements only) and an
M3046A Monitor with software Rev.E . No
measurements (for example CO
from the Measurement Extension are available.
Exchange the Measurement Server for an M3001A
with option #C06 or #C18.
The M3046A Monitor does not support some of
the measurements in a measurement extension.
Consider using a monitor that supports these
measurements.
Either
M3001A Measurement Server is connected to an
M3046A Monitor with Rev. D or lower soft-
ware. This conbination does not allow monitor-
ing. Use only M3000A Measurement Servers
with M3046A Monitors with Rev. D or lower
software.
Or
M3000A Measurement Server is connected to an
M3046A Monitor with Rev. E software. This
conbination does not allow monitoring. Use only
M3001A Measurement Servers with M3046A
Monitors with Rev. E software.
, 2nd Press/Temp)
2
Troubleshooting the Instrument